Endotracheal tube cuff pressure: need for precise measurement.
Braz, J R; Navarro, L H; Takata, I H; Nascimento Júnior, P.
Afiliación
  • Braz JR; Department of Anesthesiology, Faculdade de Medicina, Universidade Estadual Paulista, Botucatu, Brazil. jbraz@fmb.unesp.br
Sao Paulo Med J ; 117(6): 243-7, 1999 Nov 04.
Article en En | MEDLINE | ID: mdl-10625887
ABSTRACT
CONTEXT High compliance endotracheal tubes cuffs are used to prevent gas leak and also pulmonary aspiration in mechanically ventilated patients. However, the use of the usual cuff inflation volumes may cause tracheal damage.

OBJECTIVE:

We tested the hypothesis that endotracheal tube cuff pressures are routinely high (above 40 cmH2O) in the Post Anesthesia Care Unit (PACU) or Intensive Care Units (ICU).

DESIGN:

Cross-sectional study.

SETTING:

Post anesthesia care unit and intensive care unit.

PARTICIPANTS:

We measured endotracheal tubes cuff pressure in 85 adult patients, as follows G1 (n = 31) patients from the ICU; G2 (n = 32) patients from the PACU, after anesthesia with nitrous oxide; G3 (n = 22) patients from the PACU, after anesthesia without nitrous oxide. Intracuff pressure was measured using a manometer (Mallinckrodt, USA). Gas was removed as necessary to adjust cuff pressure to 30 cmH2O. MAIN MEASUREMENTS Endotracheal tube cuff pressure.

RESULTS:

High cuff pressure (> 40 cmH2O) was observed in 90.6% patients of G2, 54.8% of G1 and 45.4% of G3 (P < 0.001). The volume removed from the cuff in G2 was higher than G3 (P < 0.05).

CONCLUSION:

Endotracheal tubes cuff pressures in ICU and PACU are routinely high and significant higher when nitrous oxide is used. Endotracheal tubes cuff pressure should be routinely measured to minimize tracheal trauma.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Intubación Intratraqueal Tipo de estudio: Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Sao Paulo Med J Asunto de la revista: MEDICINA Año: 1999 Tipo del documento: Article País de afiliación: Brasil
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Intubación Intratraqueal Tipo de estudio: Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Sao Paulo Med J Asunto de la revista: MEDICINA Año: 1999 Tipo del documento: Article País de afiliación: Brasil