Your browser doesn't support javascript.
loading
Ring annuloplasty prevents delayed leaflet coaptation and mitral regurgitation during acute left ventricular ischemia.
Timek, T; Glasson, J R; Dagum, P; Green, G R; Nistal, J F; Komeda, M; Daughters, G T; Bolger, A F; Foppiano, L E; Ingels, N B; Miller, D C.
Afiliación
  • Timek T; Department of Cardiovascular and Thoracic Surgery, Stanford University School of Medicine, Stanford, CA 94305-5247, USA.
J Thorac Cardiovasc Surg ; 119(4 Pt 1): 774-83, 2000 Apr.
Article en En | MEDLINE | ID: mdl-10733769
ABSTRACT

OBJECTIVE:

Incomplete mitral leaflet coaptation during acute left ventricular ischemia is associated with end-diastolic mitral annular dilatation and ischemic mitral regurgitation. Annular rings were implanted in sheep to investigate whether annular reduction alone is sufficient to prevent mitral regurgitation during acute posterolateral left ventricular ischemia.

METHODS:

Radiopaque markers were inserted around the mitral anulus, on papillary muscle tips, and on the central meridian of both mitral leaflets in three groups of sheep control (n = 5), Physio ring (n = 5) (Baxter Cardiovascular Div, Santa Ana, Calif), and Duran ring (n = 6) (Medtronic Heart Valve Div, Minneapolis, Minn). After 8 +/- 1 days, animals were studied with biplane videofluoroscopy before and during left ventricular ischemia. Annular area was calculated from 3-dimensional marker coordinates and coaptation defined as minimal distance between leaflet edge markers.

RESULTS:

Before ischemia, leaflet coaptation occurred just after end-diastole in all groups (control 17 +/- 41, Duran 33 +/- 30, Physio 33 +/- 24 ms, mean +/- SD, P >.2 by analysis of variance). During ischemia, regurgitation was detected in all control animals, and leaflet coaptation was delayed to 88 +/- 8 ms after end-diastole (P =.02 vs preischemia). This was associated with increased end-diastolic annular area (8.0 +/- 0.9 vs 6.7 +/- 0.6 cm(2), P =.004) and septal-lateral annular diameter (2.9 +/- 0.1 vs 2.5 +/- 0.1 cm, P =.02). Mitral regurgitation did not develop in Duran or Physio sheep, time to coaptation was unchanged (Duran 25 +/- 25 ms, Physio 30 +/- 48 ms [both P >.2 vs preischemia]), and annular area remained fixed.

CONCLUSION:

Mitral annular area reduction and fixation with an annuloplasty ring eliminated delayed leaflet coaptation and prevented mitral regurgitation during acute left ventricular ischemia after ring implantation.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Isquemia Miocárdica / Implantación de Prótesis de Válvulas Cardíacas / Válvula Mitral / Insuficiencia de la Válvula Mitral Límite: Animals Idioma: En Revista: J Thorac Cardiovasc Surg Año: 2000 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Isquemia Miocárdica / Implantación de Prótesis de Válvulas Cardíacas / Válvula Mitral / Insuficiencia de la Válvula Mitral Límite: Animals Idioma: En Revista: J Thorac Cardiovasc Surg Año: 2000 Tipo del documento: Article País de afiliación: Estados Unidos