Diagnosis of the cause of malfunction of indwelling catheters for insulin injections by the use of digital fluoroscopy.
Hanas, R; Stanke, C G; Ostberg, H.
Afiliación
  • Hanas R; Department of Paediatrics, Uddevalla Hospital, Sweden. ragnar.hanas@bll.se
Pediatr Radiol ; 30(10): 674-6, 2000 Oct.
Article en En | MEDLINE | ID: mdl-11075598
ABSTRACT

BACKGROUND:

Subcutaneous indwelling catheters are used for reducing pain when injecting insulin and other drugs.

OBJECTIVE:

To use digital fluoroscopy for the diagnosis of catheter malfunction. MATERIAL AND

METHODS:

Seven children (aged 5-11 years) with diabetes mellitus were studied. They used indwelling catheters (Insuflon) for insulin injections. Contrast medium was injected into the problem catheter.

RESULTS:

The subcutaneous position of the catheter was correct in all cases, but we found two cases of delayed absorption caused by lipohypertrophy and one case of leakage.

CONCLUSIONS:

Digital fluoroscopy is a useful method for determining potential problems with indwelling catheters.