Lung volume recruitment in lambs during high-frequency oscillatory ventilation using respiratory inductive plethysmography.
Pediatr Res
; 49(1): 38-44, 2001 Jan.
Article
en En
| MEDLINE
| ID: mdl-11134489
Monitoring lung volume is important in the treatment of acute hypoxemic respiratory failure. However, there are no tools available for lung volume measurement to guide ventilator management during high-frequency oscillatory ventilation (HFOV) and during dynamic changes in conventional ventilation (CV). We studied the performance of a new respiratory inductive plethysmograph (RIP) with modified software. We measured Delta changes in lung volume above end-expiratory volume (V(RIP)) during HFOV and studied whether changes in V(RIP) parallel changes in mean airway pressure. Calibration of the plethysmograph was made by serial injections of a known gas volume in six term (140 d gestation) and eight preterm (125 d gestation) lambs. Linear regression analysis of the relationship between injected gas volume and V(RIP) showed strong correlation (r(2) = 0.93-1.00 term animals, r(2) = 0.86-1.00 preterm animals). The pressure volume curves from the calibration with the injected gas volumes also correlated well with the pressure volume curves extrapolated from changes in V(RIP). Lung hysteresis was clearly demonstrated with RIP after changes in mean airway pressure during HFOV and after changes in positive end-expiratory pressure during CV. We conclude that measurements of lung volume in term and preterm lambs by use of modified RIP correlate well with changes in mean airway pressure during HFOV, with static pressure volume curves and with changes in positive end-expiratory pressure during CV. We speculate that this technique may provide clinically useful information about changes in lung volume during HFOV and CV. However, evaluation of the precision and chronic stability of RIP measurements over prolonged periods will require further studies.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Pletismografía
/
Pulmón
Límite:
Animals
Idioma:
En
Revista:
Pediatr Res
Año:
2001
Tipo del documento:
Article
País de afiliación:
Suecia
Pais de publicación:
Estados Unidos