Your browser doesn't support javascript.
loading
Catheter tracking for MR fluoroscopy: design of a transmit/receive coil for use with a nasogastric tube.
Herlihy, D J; Larkman, D J; deSouza, N M; Williams, A D; Young, I R.
Afiliación
  • Herlihy DJ; Robert Steiner MRI Unit, ICSM (Hammersmith Campus), London W12 0HS, England, United Kingdom.
J Magn Reson Imaging ; 13(1): 127-30, 2001 Jan.
Article en En | MEDLINE | ID: mdl-11169814
ABSTRACT
A coil tuned to 21.3 MHz was incorporated into a nasogastric tube and used as a marker of tube position during magnetic resonance (MR) imaging in a 0.5-T scanner. Catheter tracking was investigated with the coil used in both transmit/receive and in receive-only modes. Data acquired from this coil were overlaid on images obtained using the body coil of the scanner. Visualization of the full length of the catheter with local high signal at the tip was achieved with a temporal resolution of approximately 1 second. J. Magn. Reson. Imaging 2001;13127-130.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Intubación Gastrointestinal Límite: Humans Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2001 Tipo del documento: Article País de afiliación: Reino Unido
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Intubación Gastrointestinal Límite: Humans Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2001 Tipo del documento: Article País de afiliación: Reino Unido