Electrode performance beyond the manufacturer's recommended lifespan using the Dornier MFL 5000 lithotripter.
Honey, R J; Weir, M J; Tariq, N.
Afiliación
  • Honey RJ; Division of Urology, Saint Michael's Hospital and University of Toronto, Canada. rjohn.honey@utoronto.ca
J Endourol ; 15(3): 243-6, 2001 Apr.
Article en En | MEDLINE | ID: mdl-11339388
ABSTRACT
BACKGROUND AND

PURPOSE:

Electrodes used by the Dornier MFL 500 lithotripter have a specified lifespan, after which, the manufacturer recommends replacing the electrode before continuing the treatment. The aim of this study was to investigate whether electrodes continue to function effectively beyond 100% consumption by measuring the pressure output of electrodes up to 300% consumption or until electrode failure. MATERIALS AND

METHODS:

We assessed new, refurbished, and twice-refurbished electrodes to compare their respective performances.

RESULTS:

Rather than a decrease, there was a trend toward increased pressure output beyond 100% electrode consumption, with no drop in pressure prior to 300% consumption or electrode failure. All three types of electrodes functioned adequately above 100% consumption.

CONCLUSION:

These observations suggest that discarding the electrode before the completion of the treatment or electrode failure is not warranted. Using a single electrode rather than multiple electrodes to complete a treatment has the potential to save both time and cost.