[Accuracy and usefulness of ultraportable hand-carried echocardiography system].
Ohyama, R; Murata, K; Tanaka, N; Takaki, A; Kimura, K; Ueda, K; Liu, J; Wada, Y; Harada, N; Matsuzaki, M.
Afiliación
  • Ohyama R; Second Department of Internal Medicine, Yamaguchi University School of Medicine, Minamikogushi 1-1-1, Ube, Yamaguchi 755-8505.
J Cardiol ; 37(5): 257-62, 2001 May.
Article en Ja | MEDLINE | ID: mdl-11392894
OBJECTIVES: The diagnostic accuracy and usefulness of an ultraportable hand-carried echocardiography system were investigated for assessing ventricular systolic function and severity of mitral valvular regurgitation. METHODS: The study population consisted of 77 consecutive patients (47 men, 30 women, mean age 63 +/- 15 years). Left ventricular end-diastolic dimension, left ventricular end-systolic dimension and left ventricular ejection fraction were measured using the hand-carried echo system and the data were compared with measurements by the conventional echocardiography system using simple linear regression analysis. Left ventricular wall motion was compared between the systems using a 16-segment model recommended by the American Society of Echocardiography. Severity of mitral regurgitation was assessed by the distance of the regurgitant signal in the left atrium. RESULTS: Left ventricular end-diastolic dimension, left ventricular end-systolic dimension and left ventricular ejection fraction showed good correlations between hand-carried and conventional echo systems (r = 0.94, 0.91 and 0.81, respectively; each p < 0.0001). The accuracy for assessing left ventricular wall motion was 94% (449 of 480 segments). The echo systems also showed the same degree of diagnostic accuracy for severity of mitral regurgitation. CONCLUSIONS: The hand-carried echo system provides accurate assessment of left ventricular function and mitral regurgitation simular to conventional echo machines.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ecocardiografía / Función Ventricular Izquierda / Insuficiencia de la Válvula Mitral / Contracción Miocárdica Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: Ja Revista: J Cardiol Asunto de la revista: CARDIOLOGIA Año: 2001 Tipo del documento: Article Pais de publicación: Países Bajos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ecocardiografía / Función Ventricular Izquierda / Insuficiencia de la Válvula Mitral / Contracción Miocárdica Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: Ja Revista: J Cardiol Asunto de la revista: CARDIOLOGIA Año: 2001 Tipo del documento: Article Pais de publicación: Países Bajos