Radioisotopic control for baclofen pump catheter failure.
Spinal Cord
; 39(5): 283-5, 2001 May.
Article
en En
| MEDLINE
| ID: mdl-11438846
ABSTRACT
STUDY DESIGN:
Case report of Baclofen pump catheter failure investigated by radioisotope injection.OBJECTIVES:
To report a safe and reliable method for evaluating catheter dysfunction.SETTING:
France.METHODS:
Single case report of failure of Baclofen pump investigated by radioisotope injection.RESULTS:
The injection demonstrated the block in the catheter. The catheter failure was not visualised by plain X-ray nor by filling the pump with radio-opaque solution.CONCLUSION:
Catheter failure is a common cause of intrathecal drug delivery problems and may be difficult to diagnose. When catheter disconnection, kink, or dislodgement is not visible on X-ray, radioisotopic control is a safe and reliable method for assessment.