Tuning compounds for electrospray ionization/in-source collision-induced dissociation and mass spectra library searching.
Weinmann, W; Stoertzel, M; Vogt, S; Svoboda, M; Schreiber, A.
Afiliación
  • Weinmann W; Institute of Legal Medicine, University Hospital, Albert Ludwigs University Freiburg, Albertstrasse 9, D-79104 Freiburg, Germany. weinmann@ukl.uni-freiburg.de
J Mass Spectrom ; 36(9): 1013-23, 2001 Sep.
Article en En | MEDLINE | ID: mdl-11599079
ABSTRACT
Tuning compounds for positive and negative electrospray ionization (ESI) were tested for the tuning of in-source collision-induced dissociation (ESI/CID) with three types of SCIEX API instruments (API 365, 2000 and 3000) in the single-quadrupole mode. The vacuum interfaces of these instruments differ slightly in geometry, but the principles of ionization and solvent evaporation by nebulizer and curtain gases, orifice and skimmer are identical. For comparison of in-source CID, breakdown curves of haloperidol, paracetamol, metronidazole and metamizole were acquired by increasing the orifice voltages. The API 2000 and 3000 required higher orifice voltages than did the API 365 to induce a similar degree of fragmentation of the protonated or deprotonated molecules to characteristic fragment ions. This increase of orifice voltage could be demonstrated with each of the four compounds tested by a shift of the maxima of the breakdown curves to higher orifice voltages. A procedure with three collision energy (CE) levels for drug identification with a mass spectra library set up with an API 365 therefore required an adjustment of the orifice voltages to higher values when being transferred to an API 2000 or API 3000. The corresponding orifice voltages for the three instruments were 20/50/80 V (API 365), 30/90/130 V (API 2000) and 40/80/120 V (API 3000). However, a change in orifice voltage of +/-10 V (with the API 2000 and 3000) hardly influenced the fit values of a library search for each single CE level. For adjusting orifice voltages with different instruments, a tuning procedure with haloperidol and paracetamol is presented. With this tuning procedure an ESI/CID mass spectra library set up for API 365 and API 150 could also be used for drug identification with an API 2000 and an API 3000 with good library search results.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Preparaciones Farmacéuticas Tipo de estudio: Prognostic_studies Idioma: En Revista: J Mass Spectrom Año: 2001 Tipo del documento: Article País de afiliación: Alemania
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Preparaciones Farmacéuticas Tipo de estudio: Prognostic_studies Idioma: En Revista: J Mass Spectrom Año: 2001 Tipo del documento: Article País de afiliación: Alemania