Flow-through chemiluminescence sensor using immobilized histamine oxidase from Arthrobacter crystallopoietes KAIT-B-007 and peroxidase for selective determination of histamine.
Anal Sci
; 17(10): 1161-4, 2001 Oct.
Article
en En
| MEDLINE
| ID: mdl-11990588
A flow sensor with immobilized oxidases is proposed for the determination of histamine in fish meat. Chemiluminometric measurement of histamine was based on the luminol reaction with hydrogen peroxide produced by immobilized histamine oxidase (EC 1.4.3.-.) and peroxidase (EC 1.11.1.7.) within a flow cell. Histamine oxidase was found in cells of Arthrobacter crystallopoietes KAIT-B-007 isolated from soil. The oxidase and peroxidase were coimmobilized covalently on tresylated hydrophilic vinyl polymer beads and packed into transparent PTFE; the tubing was used as the flow cell. One assay for histamine was done at intervals of 2 min without carryover. The calibration curve for histamine was linear from 0.1 microM to 50 microM. The response was reproducible within 1.25% of the relative standard deviation for 115-replicate injections of 50 microM histamine. The sensor system was applied to the determination of histamine in fish meat extracts.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Arthrobacter
/
Histamina
/
Amina Oxidasa (conteniendo Cobre)
/
Peroxidasa
/
Enzimas Inmovilizadas
Límite:
Animals
Idioma:
En
Revista:
Anal Sci
Año:
2001
Tipo del documento:
Article
País de afiliación:
Japón
Pais de publicación:
Suiza