Compact, low-cost Ti:Al2O3 laser for in vivo ultrahigh-resolution optical coherence tomography.
Unterhuber, A; Povazay, B; Hermann, B; Sattmann, H; Drexler, W; Yakovlev, V; Tempea, G; Schubert, C; Anger, E M; Ahnelt, P K; Stur, M; Morgan, J E; Cowey, A; Jung, G; Le, T; Stingl, A.
Afiliación
  • Unterhuber A; Department of Medical Physics, Christian Doppler Laboratory, University of Vienna, Vienna A-1090, Austria.
Opt Lett ; 28(11): 905-7, 2003 Jun 01.
Article en En | MEDLINE | ID: mdl-12816241
ABSTRACT
A compact, low-cost, prismless TiAl2O3 laser with 176-nm bandwidth (FWHM) and 20-mW output power was developed. Ultrahigh-resolution ophthalmic optical coherence tomography (OCT) ex vivo imaging in an animal model with approximately 1.2-microm axial resolution and in vivo imaging in patients with macular pathologies with approximately 3-microm axial resolution were demonstrated. Owing to the pump laser, this light source significantly reduces the cost of broadband OCT systems. Furthermore, the source has great potential for clinical application of spectroscopic and ultrahigh-resolution OCT because of its small footprint (500 mm x 180 mm including the pump laser), user friendliness, stability, and reproducibility.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía / Óptica y Fotónica / Rayos Láser Tipo de estudio: Health_economic_evaluation Límite: Animals Idioma: En Revista: Opt Lett Año: 2003 Tipo del documento: Article País de afiliación: Austria
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía / Óptica y Fotónica / Rayos Láser Tipo de estudio: Health_economic_evaluation Límite: Animals Idioma: En Revista: Opt Lett Año: 2003 Tipo del documento: Article País de afiliación: Austria