Design of a stability-detecting device for dental implants.
Huang, H M; Cheng, K Y; Chen, C F; Ou, K L; Li, C T; Lee, S Y.
Afiliación
  • Huang HM; Graduate Institute of Oral Sciences, Taipei Medical University, Taipei, Taiwan.
Proc Inst Mech Eng H ; 219(3): 203-11, 2005 May.
Article en En | MEDLINE | ID: mdl-15934396
ABSTRACT
Resonance frequency (RF) analysis technology was used to design a dental implant stability detector. The device uses a miniature-sized electromagnetic triggering rod to elicit vibration in a dental implant. Vibrational signals were recorded via an acoustic receiver. To assess the in vivo performance of the test apparatus, animal models were used. Implants were placed in the left tibia of 12 rabbits using a conventional surgical procedure. Standard 3.2 mm x 8 mm implants were placed in each test tibia with pre-tapping cavities of 3.2 mm and 3.7 mm diameters to simulate either a 'well-fitting' or a 'loosely fitting' situation. The RF values of the test implants were detected by the newly developed device which was directly mounted on the healing abutments of the implants. The results showed that the RF values of the implants under well-fitting conditions significantly increased (p < 0.01) 3 weeks after surgery and reached a plateau at around 6-7 weeks. Meanwhile implants with higher initial RF values had shorter healing times and higher final RF values at the plateau. Based on these findings, it was concluded that the idea of using the current designed device for detecting the degree of bone healing during the osseointegration process seems feasible.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Estimulación Física / Transductores / Cicatrización de Heridas / Oseointegración / Análisis de Falla de Equipo / Fracaso de la Restauración Dental / Movimiento Tipo de estudio: Evaluation_studies / Prognostic_studies Límite: Animals Idioma: En Revista: Proc Inst Mech Eng H Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2005 Tipo del documento: Article País de afiliación: Taiwán
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Estimulación Física / Transductores / Cicatrización de Heridas / Oseointegración / Análisis de Falla de Equipo / Fracaso de la Restauración Dental / Movimiento Tipo de estudio: Evaluation_studies / Prognostic_studies Límite: Animals Idioma: En Revista: Proc Inst Mech Eng H Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2005 Tipo del documento: Article País de afiliación: Taiwán
...