Your browser doesn't support javascript.
loading
[Principles of fixation of the cementless modular revision stem Revitan]. / Fixationsprinzipien des zementlosen modularen Hüftrevisionsschaftes Revitan. Eine anatomische Studie.
Fink, B; Fuerst, M; Hahn, M; Thybaud, L; Sieber, H-P; Delling, G.
Afiliación
  • Fink B; Klinik für Endoprothetik, Allgemeine und Rheumaorthopädie, Orthopädische Klinik Markgröningen.
Unfallchirurg ; 108(12): 1029-32, 1034-7, 2005 Dec.
Article en De | MEDLINE | ID: mdl-16133288
ABSTRACT
AIM AND

METHOD:

To analyse the femoral fixation of a modular cementless revision endoprosthesis, eight prosthetic combinations of the Revitan-System (Centerpulse, Winterthur, Switzerland) were implanted in four cadavers. On three cadavers a curved revision Revitan stem was implanted by an endofemoral approach on one side and by a transfemoral approach on the contralateral side. On the fourth cadaver a straight Revitan stem was implanted on one side and a curved Revitan stem on the contralateral side using a transfemoral approach. Transversal slides of 7-8 mm thickness were performed at the fixation areas of each implant and the implant-bone contact was analysed macroscopically and using contact radiography.

RESULTS:

The straight stem implanted by a transfemoral approach showed a double-conical press-fit fixation with cutting of the eight longitudinal fins into the cortical bone. The curved revision stems implanted by the same approach had a circular surface fixation similar to the press-fit fixation of the straight stem. In contrast, the curved stems implanted by the endofemoral approach (without a window) showed a three-surface fixation. Hereby the two distal fixation areas led to the primary implant stability by three of the four double edges of the octagonal cross-sectional area cutting into the cortical bone. At the proximal fixation zone the implant only had contact of two implant double edges to the cortical bone.

CONCLUSION:

Different approaches for implantation lead to different fixation techniques of a curved revision stem. This should be considered by analysing postoperative sintering rates of cementless revision stems.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Prótesis de Cadera Límite: Aged / Female / Humans / Male Idioma: De Revista: Unfallchirurg Asunto de la revista: TRAUMATOLOGIA Año: 2005 Tipo del documento: Article
Buscar en Google