Development and application of programs to measure modulation transfer function, noise power spectrum and detective quantum efficiency.
Padgett, R; Kotre, C J.
Afiliación
  • Padgett R; Regional Medical Physics Department, Newcastle General Hospital, Newcastle upon Tyne NE4 6BE, UK. Rodney.Padgett@nuth.northy.nhs.uk
Radiat Prot Dosimetry ; 117(1-3): 283-7, 2005.
Article en En | MEDLINE | ID: mdl-16461517
ABSTRACT
This project aimed to produce programs to calculate the modulation transfer function (MTF), noise power spectrum (NPS) and detective quantum efficiency (DQE) of digital X-ray systems, given a suitable digital image. The MTF was calculated using the edge technique and the NPS was calculated from a flat field image. Both programs require a suitably edited DICOM image as input. The DQE was then calculated from the output of MTF and NPS programs. This required data external to the DQE program to estimate the number of quanta per mm2 in the beam which formed the NPS image. All three programs run independent of each other on a PC and require no special software to be installed. Results for MTF, NPS and DQE for a Philips AC3 CR system are presented. In addition, the results for MTF from a Siemens Duo CT scanner with a specially designed PTFE edge are also shown.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Protección Radiológica / Radiometría / Intensificación de Imagen Radiográfica Aspecto: Implementation_research Límite: Humans Idioma: En Revista: Radiat Prot Dosimetry Año: 2005 Tipo del documento: Article País de afiliación: Reino Unido
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Protección Radiológica / Radiometría / Intensificación de Imagen Radiográfica Aspecto: Implementation_research Límite: Humans Idioma: En Revista: Radiat Prot Dosimetry Año: 2005 Tipo del documento: Article País de afiliación: Reino Unido
...