Chronic peritoneal dialysis catheters: challenges and design solutions.
Int J Artif Organs
; 29(1): 85-94, 2006 Jan.
Article
en En
| MEDLINE
| ID: mdl-16485243
Although highly successful as transcutaneous access devices, today's peritoneal dialysis catheters still have imperfect hydraulic function, biocompatibility and resistance to infection. Success of Tenckhoff catheters is greatly improved by the proper positioning of deep and subcutaneous cuffs and intraperitoneal segment. Newer peritoneal catheter designs are intended to improve hydraulic function, avoid outflow failure, and diminish exit site infection. These catheter designs serve as excellent alternatives for patients with various types of failure of Tenckhoff catheters. Catheters have been designed for Continuous Flow Peritoneal Dialysis, and have generally been successful in providing high peritoneal dialysis flow rate, but not always successful in optimally distributing flow of peritoneal fluid. Improvements in catheter design may expand the use of peritoneal dialysis as a successful home dialysis therapy.