[Radiation synovectomy of the knee joint: Evaluation of bremsstrahlung-detection by using a corpse phantom]. / Radiosynoviorthese des Kniegelenkes: Phantomuntersuchung zur Bremsstrahlenmessung.
Klett, R; Puille, M; Steiner, D; Bauer, R.
Afiliación
  • Klett R; Klinik für Nuklearmedizin, Universitätsklinikum Giessen und Marburg, Friedrichstrasse 25, 35385 Giessen. rigobert.klett@radiol.med.uni-giessen.de
Nuklearmedizin ; 45(1): 57-61, 2006.
Article en De | MEDLINE | ID: mdl-16493515
ABSTRACT

AIM:

Multiple procedures for the quantification of activity leakage in radiation synovectomy of the knee joint have been described in the literature. We compared these procedures considering the real conditions of dispersion and absorption using a corpse phantom.

METHODS:

We simulated different distributions of the activity in the knee joint and a different extra-articular spread into the inguinal lymph nodes. The activity was measured with a gamma-camera. Activity leakage was calculated by measuring the retention in the knee joint only using an anterior view, using the geometric mean of anterior and posterior views, or using the sum of anterior and posterior views. The same procedures were used to quantify the activity leakage by measuring the activity spread into the inguinal lymph nodes. In addition, the influence of scattered rays was evaluated.

RESULTS:

For several procedures we found an excellent association with the real activity leakage, shown by an r(2) between 0.97 and 0.98. When the real value of the leakage is needed, e. g. in dosimetric studies, simultaneously measuring of knee activity and activity in the inguinal lymph nodes in anterior and posterior views and calculation of the geometric mean with exclusion of the scatter rays was found to be the procedure of choice.

CONCLUSION:

When measuring of activity leakage is used for dosimetric calculations, the above-described procedure should be used. When the real value of the leakage is not necessary, e. g. for comparing different therapeutic modalities, several of the procedures can be considered as being equivalent.