Outbreak of toxic anterior segment syndrome following cataract surgery associated with impurities in autoclave steam moisture.
Infect Control Hosp Epidemiol ; 27(3): 294-8, 2006 Mar.
Article en En | MEDLINE | ID: mdl-16532418
ABSTRACT

BACKGROUND:

Toxic anterior segment syndrome (TASS), a complication of cataract surgery, is a sterile inflammation of the anterior chamber of the eye. An outbreak of TASS was recognized at an outpatient surgical center and its affiliated hospital in December 2002.

METHODS:

Medical records of patients who underwent cataract surgery during the outbreak were reviewed, and surgical team members who participated in the operations were interviewed. Potential causes of TASS were identified and eliminated. Feedwater from autoclave steam generators and steam condensates were analyzed by use of spectroscopy and ion chromatography.

RESULTS:

During the outbreak, 8 (38%) of 21 cataract operations were complicated by TASS, compared with 2 (0.07%) of 2,713 operations performed from January 1996 through November 2002. Results of an initial investigation suggested that cataract surgical equipment may have been contaminated by suboptimal equipment reprocessing or as a result of personnel changes. The frequency of TASS decreased (1 of 44 cataract operations) after reassignment of personnel and revision of equipment reprocessing procedures. Further investigation identified the presence of impurities (eg, sulfates, copper, zinc, nickel, and silica) in autoclave steam moisture, which was attributed to improper maintenance of the autoclave steam generator in the outpatient surgical center. When impurities in autoclave steam moisture were eliminated, no cases of TASS were observed after more than 1,000 cataract operations.

CONCLUSION:

Suboptimal reprocessing of cataract surgical equipment may evolve over time in busy, multidisciplinary surgical centers. Clinically significant contamination of surgical equipment may result from inappropriate maintenance of steam sterilization systems. Standardization of protocols for reprocessing of cataract surgical equipment may prevent outbreaks of TASS and may be of assistance during outbreak investigations.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Contaminación del Agua / Extracción de Catarata / Contaminación de Equipos / Brotes de Enfermedades / Inflamación Tipo de estudio: Etiology_studies / Guideline / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Infect Control Hosp Epidemiol Asunto de la revista: DOENCAS TRANSMISSIVEIS / ENFERMAGEM / EPIDEMIOLOGIA / HOSPITAIS Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Contaminación del Agua / Extracción de Catarata / Contaminación de Equipos / Brotes de Enfermedades / Inflamación Tipo de estudio: Etiology_studies / Guideline / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Infect Control Hosp Epidemiol Asunto de la revista: DOENCAS TRANSMISSIVEIS / ENFERMAGEM / EPIDEMIOLOGIA / HOSPITAIS Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos
...