Your browser doesn't support javascript.
loading
Performance of the electronic personal dosemeter for neutron 'Saphydose-N' at different workplaces of nuclear facilities.
Lahaye, T; Chau, Q; Ménard, S; Lacoste, V; Muller, H; Luszik-Bhadra, M; Reginatto, M; Bruguier, P.
Afiliación
  • Lahaye T; Institute for Radiological Protection and Nuclear Safety, BP 17, 92262 Fontenay-aux-Roses, France.
Radiat Prot Dosimetry ; 120(1-4): 383-6, 2006.
Article en En | MEDLINE | ID: mdl-16820401
ABSTRACT
This paper mainly aims at presenting the measurements and the results obtained with the electronic personal neutron dosemeter Saphydose-N at different facilities. Three campaigns were led in the frame of the European contract EVIDOS ('Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields'). The first one consisted in the measurements at the IRSN French research laboratory in reference neutron fields generated by a thermal facility (SIGMA), radionuclide ISO sources ((241)AmBe; (252)Cf; (252)Cf(D(2)O)\Cd) and a realistic spectrum (CANEL/T400). The second one was performed at the Krümmel Nuclear Power Plant (Germany) close to the boiling water reactor and to a spent fuel transport cask. The third one was realised at Mol (Belgium), at the VENUS Research Reactor and at Belgonucléaire, a fuel processing factory.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Protección Radiológica / Dosimetría Termoluminiscente / Exposición Profesional / Neutrones / Reactores Nucleares Tipo de estudio: Diagnostic_studies / Evaluation_studies País/Región como asunto: Europa Idioma: En Revista: Radiat Prot Dosimetry Año: 2006 Tipo del documento: Article País de afiliación: Francia
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Protección Radiológica / Dosimetría Termoluminiscente / Exposición Profesional / Neutrones / Reactores Nucleares Tipo de estudio: Diagnostic_studies / Evaluation_studies País/Región como asunto: Europa Idioma: En Revista: Radiat Prot Dosimetry Año: 2006 Tipo del documento: Article País de afiliación: Francia