Your browser doesn't support javascript.
loading
Derivation of V function for LR 115 SSNTD from its sensitivity to 220Rn in a diffusion chamber.
Leung, S Y Y; Nikezic, D; Leung, J K C; Yu, K N.
Afiliación
  • Leung SY; Department of Physics and Materials Science, City University of Hong Kong, Tat Chee Avenue, Kowloon Tong, Kowloon, Hong Kong.
Appl Radiat Isot ; 65(3): 313-7, 2007 Mar.
Article en En | MEDLINE | ID: mdl-17029818
ABSTRACT
The sensitivity of the LR 115 detector inside a diffusion chamber to (220)Rn gas concentration is dependent on the removed active layer thickness during chemical etching. This dependence is related to the V function for the LR 115 detector (where V is the ratio between the track etch velocity V(t) and the bulk etch velocity V(b)) and the geometry of the diffusion chamber. The present paper presents the experimentally determined relationship between the sensitivity of the LR 115 detector inside a Karlsruhe diffusion chamber (determined from the number of etched tracks completely penetrating the active cellulose nitrate layer) and the removed active layer thickness. These data were used to derive the V function for the LR 115 detector, which took the functional form of the Durrani-Green's function, i.e., V=1+((a1e-)(a2R+a3e(-a)4R))(1(-e)(-a5R)), with the best-fitted constants as a(1)=14.50, a(2)=0.50, a(3)=3.9 and a(4)=0.066.
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Diagnostic_studies Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2007 Tipo del documento: Article País de afiliación: Hong Kong
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Diagnostic_studies Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2007 Tipo del documento: Article País de afiliación: Hong Kong