Differential pulse cathodic voltammetric determination of floctafenine and metopimazine.
J Pharm Biomed Anal
; 43(4): 1535-9, 2007 Mar 12.
Article
en En
| MEDLINE
| ID: mdl-17161576
A simple, rapid and sensitive voltammetric method for the determination of floctafenine (FFN) and metopimazine (MPZ) was developed. Well-defined cathodic waves were obtained for both drugs in Britton-Robinson buffer pH 9.0 using the differential-pulse mode at the hanging mercury drop electrode (HMDE). The current-concentration relationship was found to be linear over the ranges 0.4-3.6 and 0.4-2.4 microg ml(-1) for FFN and MPZ, respectively. The quantification of the two drugs in their pharmaceutical formulations was carried out using the proposed voltammetric method and compared with spectrophotometric analysis data. The mechanisms of the electrode reactions for the two drugs were proposed.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Antiinflamatorios no Esteroideos
/
Ortoaminobenzoatos
/
Ácidos Isonipecóticos
/
Antieméticos
Tipo de estudio:
Diagnostic_studies
Idioma:
En
Revista:
J Pharm Biomed Anal
Año:
2007
Tipo del documento:
Article
País de afiliación:
Egipto
Pais de publicación:
Reino Unido