[Reoperation due to failure of a Freestyle bioprosthesis].
Kyobu Geka
; 61(7): 545-8, 2008 Jul.
Article
en Ja
| MEDLINE
| ID: mdl-18616098
ABSTRACT
We report re-do aortic valve and ascending aorta replacements by using the valve-on-valve technique for primary tissue failure of a Freestyle bioprosthesis. A 74-year-old male, who had had a 25 mm Freestyle bioprosthetic valve implanted by the sub-coronary method 5 years previously for aortic valve regurgitation due to congenital bicuspid valve, was referred to our hospital for dyspnea and palpitation. He presented with heart failure secondary to aortic regurgitation due to primary tissue failure, and computed tomography demonstrated an enlarged ascending aorta (5 cm in diameter). The operative findings revealed that the Freestyle bioprosthetic valve had a leaflet tear at the left coronary cusp. We replaced the degenerated Freestyle bioprosthesis with a 19 mm Mosaic aortic bioprosthesis by using the valve-on-valve technique, and ascending aorta replacement was performed simultaneously. This technique can be useful for re-do surgery for degenerated stentless valves to avoid potential risks of complete excision of the bioprosthesis.