Adjusting mobility scales of ion mobility spectrometers using 2,6-DtBP as a reference compound.
Viitanen, A-K; Mauriala, T; Mattila, T; Adamov, A; Pedersen, C S; Mäkelä, J M; Marjamäki, M; Sysoev, A; Keskinen, J; Kotiaho, T.
Afiliación
  • Viitanen AK; Aerosol Physics Laboratory, Department of Physics, Tampere University of Technology, P.O. Box 692, FI-33101 Tampere, Finland.
Talanta ; 76(5): 1218-23, 2008 Sep 15.
Article en En | MEDLINE | ID: mdl-18761181
ABSTRACT
Performance of several time-of-flight (TOF) type ion mobility spectrometers (IMS) was compared in a joint measurement campaign and their mobility scales were adjusted based on the measurements. A standard reference compound 2,6-di-tert butylpyridine (2,6-DtBP) was used to create a single peak ion mobility distribution with a known mobility value. The effective length of the drift tube of each device, considered here as an instrument constant, was determined based on the measurements. Sequentially, two multi-peaked test compounds, DMMP and DIMP, were used to verify the performance of the adjustment procedure in a wider mobility scale. By determining the effective drift tube lengths using 2,6-DtBP, agreement between the devices was achieved. The determination of effective drift tube lengths according to standard reference compound was found to be a good method for instrument inter-comparison. The comparison procedure, its benefits and shortcomings as well as dependency on accuracy of literature value are discussed along with the results.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Talanta Año: 2008 Tipo del documento: Article País de afiliación: Finlandia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Talanta Año: 2008 Tipo del documento: Article País de afiliación: Finlandia