Your browser doesn't support javascript.
loading
Aberration-corrected microscopy for structural biology applications.
Ziegler, A; Rockel, B; Hegerl, R; Freitag, B; Lücken, U; Plitzko, J M.
Afiliación
  • Ziegler A; Max-Planck Institute of Biochemistry, Molecular Structural Biology, D-82152 Martinsried, Germany. ziegler@biochem.mpg.de
J Microsc ; 233(1): 170-7, 2009 Jan.
Article en En | MEDLINE | ID: mdl-19196423
ABSTRACT
This study explores the potential of a C(s)-corrected transmission electron microscope for structural studies of biological samples, in particular isolated macromolecular complexes. A 300-kV transmission electron microscope, equipped with a C(s) corrector was employed to record sets of images at different defocus and C(s) settings. The experiments were designed to determine whether imaging with large defocus benefits from C(s) correction. Defocus contrast in biological imaging has a stronger influence on image resolution than any other parameter. We find the results are in good agreement with theoretical framework, verifying that the typical imaging conditions required for biological investigations are not affected by C(s) correction.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Microscopía Electrónica de Transmisión Idioma: En Revista: J Microsc Año: 2009 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Microscopía Electrónica de Transmisión Idioma: En Revista: J Microsc Año: 2009 Tipo del documento: Article País de afiliación: Alemania