Relaxation of forces needed to distract cervical vertebrae after discectomy: a biomechanical study.
Aryan, Henry E; Newman, C Benjamin; Lu, Daniel C; Hu, Serena S; Tay, Bobby K; Bradford, David S; Puttlitz, Christian M; Ames, Christopher P.
Afiliación
  • Aryan HE; Department of Neurological Surgery, University of California, San Francisco, USA.
J Spinal Disord Tech ; 22(2): 100-4, 2009 Apr.
Article en En | MEDLINE | ID: mdl-19342931
ABSTRACT
STUDY

DESIGN:

In vitro and in vivo biomechanical stress measurements are made of the intervertebral disc segment distraction force during anterior cervical discectomy.

OBJECTIVE:

The purpose of this study is to determine the short-term force relaxation of the native intervertebral disc segment and to determine the short-term force relaxation of the segment after removal of the intervertebral disc, as is commonly performed in anterior cervical discectomy with fusion and arthroplasty. SUMMARY OF BACKGROUND DATA No published data examine the issue of intraoperative distraction force of the cervical intervertebral disc segment. This is a novel research in this area.

METHODS:

In vitro and in vivo studies under institutional review board approval were performed to determine the mechanical behavior of the normal and diseased cervical functional spinal unit. Seven in vitro and 11 in vivo spines were studied. Strain measurements between distracting Caspar-type pins were made before, at various points during, and after discectomy to assess how removal of the disc and other spinal components affects the force-displacement behavior of the spinal unit.

RESULTS:

The in vitro data show progressive reduction in force needed for distraction after discectomy and uncovertebral joint resection. Greatest reduction is noted after discectomy. The in vivo data indicate that, on average, the cervical functional spinal unit requires 20 N less force to achieve the same degree of distraction after removal of the intervertebral disc.

CONCLUSIONS:

A sharp reduction in the strain across the intervertebral space occurs after distraction. The removal of the cervical intervertebral disc significantly reduces the viscoelastic response of the cervical motion segment. The long-term force used to stabilize intervertebral grafts or implants is less than what is achieved at the time of distraction. The exact magnitude of the resultant force on graft or device at a given distraction force is unknown and would depend also upon fit.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fusión Vertebral / Tracción / Vértebras Cervicales / Discectomía / Disco Intervertebral / Desplazamiento del Disco Intervertebral Tipo de estudio: Etiology_studies Límite: Humans Idioma: En Revista: J Spinal Disord Tech Asunto de la revista: ORTOPEDIA Año: 2009 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fusión Vertebral / Tracción / Vértebras Cervicales / Discectomía / Disco Intervertebral / Desplazamiento del Disco Intervertebral Tipo de estudio: Etiology_studies Límite: Humans Idioma: En Revista: J Spinal Disord Tech Asunto de la revista: ORTOPEDIA Año: 2009 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA