IFIN-HH ionization chamber calibration and its validation; electrometric system improvement.
Appl Radiat Isot
; 68(7-8): 1266-9, 2010.
Article
en En
| MEDLINE
| ID: mdl-20004588
The paper presents the results obtained in the calibration of the CENTRONIC IG12/20A ionization chamber for 18 gamma ray emitters, and its improvement with a Keithley 6517A electrometer. The calibration figures were determined either directly in pA MBq(-1) units, or calculated from old units, by using the system capacity value. The calibration figures, determined with RML's standards, are compared with those deduced from the KCRV or the mean of the comparisons, and the values determined at PTB-Germany.