Anaesthetic gas monitoring comparison between two side-stream monitors.
Heijbel, H; Ostblom, J; Brattwall, M; Jakobsson, Jan.
Afiliación
  • Heijbel H; Foot & Ankle Surgical Centre, Stockholm, Sweden. Jan.Jakobsson@KI.SE
J Clin Monit Comput ; 24(2): 169-72, 2010 Apr.
Article en En | MEDLINE | ID: mdl-20238153
ABSTRACT
End-tidal gas monitoring has become standard of care during inhaled general anaesthesia. We studied the performance of a new side stream gas monitor the ISA multi-gas monitor. The performance was studied at constant low flow of calibration gas and end-tidal anaesthetic measure was studied during routine day case anaesthesia. Pair wise readings of end-tidal halogenated anaesthetic concentration were recorded during low flow anaesthesia. Performance was found to be high; all calibration gas measures were within 0.1 vol% deviation. During routine anaesthesia mean bias was -0.036 vol% and 93 out of 97 pair-wise readings were within the agreement limits as compared to the reference Datex instrument.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Anestesia por Circuito Cerrado / Anestesia por Inhalación / Anestésicos Tipo de estudio: Diagnostic_studies / Evaluation_studies Idioma: En Revista: J Clin Monit Comput Asunto de la revista: INFORMATICA MEDICA / MEDICINA Año: 2010 Tipo del documento: Article País de afiliación: Suecia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Anestesia por Circuito Cerrado / Anestesia por Inhalación / Anestésicos Tipo de estudio: Diagnostic_studies / Evaluation_studies Idioma: En Revista: J Clin Monit Comput Asunto de la revista: INFORMATICA MEDICA / MEDICINA Año: 2010 Tipo del documento: Article País de afiliación: Suecia