Intrathoracic impedance preceding ventricular tachyarrhythmia episodes.
Moore, Hans J; Peters, Matthew N; Franz, Michael R; Karasik, Pamela E; Singh, Steven N; Fletcher, Ross D.
Afiliación
  • Moore HJ; Washington Veterans Affairs Medical Center, Washington, DC, USA. Hans.Moore@VA.Gov
Pacing Clin Electrophysiol ; 33(8): 960-6, 2010 Aug.
Article en En | MEDLINE | ID: mdl-20370851
ABSTRACT

BACKGROUND:

Heart failure is associated with ventricular tachyarrhythmias (VT/VF). Fluid accumulation during worsened heart failure may trigger VT/VF. Increased intrathoracic impedance has been correlated with fluid accumulation during heart failure. Implanted defibrillators capable of daily measures of intrathoracic impedance allow correlation of impedance with occurrence of VT/VF. We hypothesized that VT/VF episodes are preceded by decreases in intrathoracic impedance. The goal was to identify the relationship of intrathoracic impedance measured by implanted cardioverter defibrillators to the occurrence of VT/VF.

METHOD:

Implanted defibrillator follow-up data were obtained retrospectively. Those with Medtronic OptiVol (Medtronic Inc., Minneapolis, MN, USA), storing averaged daily and reference impedance values, were reviewed for VT/VF episodes. Impedance changes in the week leading up to VT/VF were analyzed.

RESULTS:

A total of 317 VT/VF episodes in a cohort of 121 patients' follow-up data were evaluated. Averaged daily intrathoracic impedance declined preceding 64% of VT/VF episodes, with an average decline of 0.46 +/- 0.35 Ohms from the day before the VT/VF episodes. However, the mean values of the averaged daily and reference impedance did not change significantly. A novel measure, DeltaTI, the sum of the daily differences between the averaged daily and reference impedance, was negative preceding 66% of VT/VF episodes (P < 0.001). The mean DeltaTI was -4.0 +/- 1.3 Ohms, which was significantly lower than the theoretically expected value of zero Ohms (P < 0.01).

CONCLUSION:

(1) Averaged daily impedance declined preceding 64% of VT/VF episodes, but the overall decline was of small magnitude; (2) a novel measure, DeltaTI, was negative preceding 66% of VT/VF episodes, and significantly below zero.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cardiografía de Impedancia / Taquicardia Ventricular / Desfibriladores Implantables Tipo de estudio: Diagnostic_studies / Etiology_studies / Prognostic_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cardiografía de Impedancia / Taquicardia Ventricular / Desfibriladores Implantables Tipo de estudio: Diagnostic_studies / Etiology_studies / Prognostic_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos