Spectrometer wavelength calibration using spectrally resolved white-light interferometry.
Youngquist, Robert C; Simmons, Stephen M; Belanger, Andrea M.
Afiliación
  • Youngquist RC; Kennedy Space Center Applied Physics Lab, Florida 32899, USA. Robert.C.Youngquist@nasa.gov
Opt Lett ; 35(13): 2257-9, 2010 Jul 01.
Article en En | MEDLINE | ID: mdl-20596212
ABSTRACT
A method is presented for calibrating the wavelength assignments of a spectrometer. White light is launched through an unbalanced Michelson interferometer to produce an optical signal that varies strongly, but predictably, with wavelength. The spectrometer output is then compared to the predicted signal and the deviations found are assumed to be due to wavelength assignment errors, permitting the spectrometer to be calibrated. The details of this process are presented with an example.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: Opt Lett Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos