Direct-injection HPLC method of measuring micafungin in human plasma using a novel hydrophobic/hydrophilic hybrid ODS column.
J Chromatogr B Analyt Technol Biomed Life Sci
; 879(13-14): 1029-32, 2011 Apr 15.
Article
en En
| MEDLINE
| ID: mdl-21439916
ABSTRACT
A direct-injection HPLC-based method has been developed for determining amounts of micafungin in human plasma using a novel hydrophobic/hydrophilic hybrid ODS column. The method is easy to perform and requires only 10 µL of a filtered plasma sample. The chromatographic separations were carried out with a gradient mode. The fluorescence detection wavelengths of excitation and emission were set at 273 nm and 464 nm, respectively. Retention times for micafungin and IS were 22.4 and 23.7 min, respectively. Micafungin and FR195743 (IS) peaks were completely separated with little tailing, and no interference was observed. The calibration curve of micafungin showed good linearity in the range of 0.5-20.0 µg/mL (r(2)=1.00). The intra-day accuracy ranged from -4.5 to 5.3%. The inter-day accuracy ranged from -9.8 to 1.5%. The precisions were less than 10%. This method is useful for the determination of micafungin in human plasma.