Local specific absorption rate in high-pass birdcage and transverse electromagnetic body coils for multiple human body models in clinical landmark positions at 3T.
J Magn Reson Imaging
; 33(5): 1209-17, 2011 May.
Article
en En
| MEDLINE
| ID: mdl-21509880
ABSTRACT
PURPOSE:
To use electromagnetic (EM) simulations to study the effects of body type, landmark position, and radiofrequency (RF) body coil type on peak local specific absorption rate (SAR) in 3T magnetic resonance imaging (MRI). MATERIALS ANDMETHODS:
Numerically computed peak local SAR for four human body models (HBMs) in three landmark positions (head, heart, pelvic) were compared for a high-pass birdcage and a transverse electromagnetic 3T body coil. Local SAR values were normalized to the IEC whole-body average SAR limit of 2.0 W/kg for normal scan mode.RESULTS:
Local SAR distributions were highly variable. Consistent with previous reports, the peak local SAR values generally occurred in the neck-shoulder area, near rungs, or between tissues of greatly differing electrical properties. The HBM type significantly influenced the peak local SAR, with stockier HBMs, extending extremities towards rungs, displaying the highest SAR. There was also a trend for higher peak SAR in the head-centric and heart-centric positions. The impact of the coil types studied was not statistically significant.CONCLUSION:
The large variability in peak local SAR indicates the need to include more than one HBM or landmark position when evaluating safety of body coils. It is recommended that an HBM with arms near the rungs be included to create physically realizable high-SAR scenarios.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Imagen por Resonancia Magnética
/
Imagen de Cuerpo Entero
Tipo de estudio:
Prognostic_studies
/
Risk_factors_studies
Límite:
Female
/
Humans
/
Male
Idioma:
En
Revista:
J Magn Reson Imaging
Asunto de la revista:
DIAGNOSTICO POR IMAGEM
Año:
2011
Tipo del documento:
Article
País de afiliación:
Estados Unidos