Refinement of ERCP by using the Olympus V-scope system with a 0.025 in. compatible and complete fixable Visiglide(®) guidewire.
Raithel, Martin; Naegel, Andreas; Seidel, Sebastian; Raithel, Sandra; Diebel, Hiwot; Neurath, Markus F; Maiss, Juergen.
Afiliación
  • Raithel M; Department of Medicine 1, University Erlangen - Nuremberg, Erlangen, Germany. martin.raithel@uk-erlangen.de
Dig Liver Dis ; 43(10): 788-91, 2011 Oct.
Article en En | MEDLINE | ID: mdl-21676663
ABSTRACT

AIM:

Prospective evaluation of the new 0.025 in. Visiglide(®) guidewire to facilitate endoscopic retrograde cholangiopancreaticography using the Olympus V-scope. MATERIALS AND

METHODS:

Interventional endoscopic retrograde cholangiopancreaticography was performed in 9 patients with the Olympus V-scope and the 0.025 in. Visiglide(®) guidewire (VS group), whilst 9 other patients underwent endoscopic retrograde cholangiopancreaticography with a conventional Olympus duodenoscope using 0.035 in. conventional guidewires (controls). Exchange time of accessories, X-ray time, dose and endoscopic retrograde cholangiopancreaticography examination time were investigated.

RESULTS:

The VS group showed a significantly lower exchange time of endoscopic retrograde cholangiopancreaticography accessories (9; 4-10s, p<0.0001) than controls (29; 19-44s). The Visiglide(®) guidewire was complete fixable by the elevator in 35/36 instrument exchanges (97%) compared to 16/31 exchanges (52%) using conventional guidewires.

LIMITATIONS:

Single-centre study, small patient numbers, two investigators.

CONCLUSIONS:

Endoscopic retrograde cholangiopancreaticography using the Olympus V-scope with the new 0.025 in. Visiglide(®) guidewire enables a 3-fold faster exchange of accessories due to a nearly complete fixation of the guidewire.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Colangiopancreatografia Retrógrada Endoscópica / Duodenoscopios Tipo de estudio: Clinical_trials / Observational_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Dig Liver Dis Asunto de la revista: GASTROENTEROLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Colangiopancreatografia Retrógrada Endoscópica / Duodenoscopios Tipo de estudio: Clinical_trials / Observational_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Dig Liver Dis Asunto de la revista: GASTROENTEROLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Alemania