Note: binding energy scale calibration of electron spectrometers for photoelectron spectroscopy using a single sample.
Helander, M G; Greiner, M T; Wang, Z B; Lu, Z H.
Afiliación
  • Helander MG; Department of Materials Science and Engineering, University of Toronto, 184 College St., Toronto, Ontario, Canada M5S 3E4. michael.helander@utoronto.ca
Rev Sci Instrum ; 82(9): 096107, 2011 Sep.
Article en En | MEDLINE | ID: mdl-21974634
ABSTRACT
The feasibility of calibrating the binding energy scale for photoelectron spectroscopy using a single sample without sputter cleaning was investigated. By measuring the Fermi level of a Au film with both monochromatic Al Kα (hν = 1486.7 eV) and He Iα (hν = 21.22 eV), the binding energy scale was simply and accurately calibrated. This method is found to yield binding energy values for the Cu 2p(3/2), Ag 3d(5/2), and Au 4f(7/2) peaks that agree with the standard tabulated values defined in International Organization for Standardization 15472 to within ±0.02 eV.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: Rev Sci Instrum Año: 2011 Tipo del documento: Article