Delivery of helium­oxygen mixture during spontaneous breathing: evaluation of three high-concentration face masks.
Roche-Campo, Ferran; Vignaux, Laurence; Galia, Fabrice; Lyazidi, Aissam; Vargas, Frédéric; Texereau, Joëlle; Apiou-Sbirlea, Gabriela; Jolliet, Philippe; Brochard, Laurent.
Afiliación
  • Roche-Campo F; Medical Intensive Care Unit, AP-HP Groupe Hospitalier Henri Mondor, Créteil, France. ferranroche@yahoo.es
Intensive Care Med ; 37(11): 1787-92, 2011 Nov.
Article en En | MEDLINE | ID: mdl-21976185
ABSTRACT

PURPOSE:

To evaluate the efficacy of delivering a mixture of helium and oxygen gas (He­O2) in spontaneous ventilation. Three high oxygen flow reservoir masks were tested the Heliox21, specifically designed for helium; the Hi-Ox80 mask, with an inspiratory and an expiratory valve; and a standard high-concentration face mask.

METHODS:

This prospective randomized crossover study was performed in six healthy volunteers in a laboratory setting. Volunteers breathed a mixture of 78% He/22% O2 through each of the masks under two different breathing conditions (rest and hyperventilation minute ventilation of 14.9 ± 6.1 and 26.7 ± 8.7 L min(−1), respectively) and four different He­O2 flow rates (7, 10, 12, and 15 L min(−1)).

RESULTS:

A nasopharyngeal catheter was used to estimate He pharyngeal concentration (Fp [He]) in the airways in order to determine the percentage of contamination with room air (% air cont) at end-expiration. Under all testing conditions, the Hi-Ox80 mask presented a significantly lower % air cont. During resting breathing pattern, a Fp [He] higher than 50% was achieved in 54% of the tests performed with the Hi-Ox80 mask compared to 29% for the Heliox21 mask and only 17% for the standard mask. At hyperventilation, a Fp [He] higher than 50% was achieved in 17% of the tests performed with the Hi-Ox mask compared to 4% for the other two masks.

CONCLUSION:

He­O2 administration via the usual high-concentration reservoir masks results in significant dilution by room air. The Hi-Ox80 mask minimized room air contamination and much more frequently achieved a pharyngeal He concentration higher than 50%.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Oxígeno / Respiración / Helio / Máscaras Tipo de estudio: Clinical_trials / Observational_studies Límite: Adult / Female / Humans / Male País/Región como asunto: Europa Idioma: En Revista: Intensive Care Med Año: 2011 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Oxígeno / Respiración / Helio / Máscaras Tipo de estudio: Clinical_trials / Observational_studies Límite: Adult / Female / Humans / Male País/Región como asunto: Europa Idioma: En Revista: Intensive Care Med Año: 2011 Tipo del documento: Article País de afiliación: Francia