Second-generation annealed highly cross-linked polyethylene exhibits low wear.
Clin Orthop Relat Res
; 470(6): 1696-704, 2012 Jun.
Article
en En
| MEDLINE
| ID: mdl-22161120
ABSTRACT
BACKGROUND:
Compared with conventional polyethylene, first-generation highly cross-linked polyethylenes have low wear, but controversy exists regarding their reduced mechanical strength and/or retained free radicals. Second-generation highly cross-linked polyethylenes have been developed to reduce wear, maintain mechanical strength, and have oxidative resistance, but it is unclear whether they do so. QUESTIONS/PURPOSES:
The primary objective of this study therefore was to determine if a second-generation annealed material has low linear wear at 5 years followup. Secondary objectives were to evaluate for overall survivorship, implant fixation, osteolysis, and effect of socket inclination on wear.METHODS:
In a multicenter prospective study, we radiographically evaluated 155 patients (167 hips) at 3 years, 124 patients (132 hips) at 4 years, and 46 patients (51 hips) at 5 years. The linear head penetration rate was measured at 6 weeks, 1 year, and yearly through 5 years.RESULTS:
The head penetration per year after the first year of bedding-in was 0.024 mm per year at 3 years, 0.020 mm per year at 4 years, and 0.008 mm per year at 5 years. The average wear rate over 5 years was 0.015 mm per year and represents a 58% improvement over a first-generation annealed highly cross-linked polyethylene. The Kaplan-Meier survivorship (revision for any reason) was 97.8%. We revised no hip for bearing surface failure and observed no osteolysis. Socket inclination did not affect linear wear.CONCLUSIONS:
These data suggest the linear wear rate for a second-generation annealed highly cross-linked polyethylene is no greater than that for historic controls of first-generation highly cross-linked polyethylenes, and no untoward complications were encountered with this new material. LEVEL OF EVIDENCE Level II, prognostic study. See the Guidelines for Authors for a complete description of levels of evidence.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Falla de Prótesis
/
Artroplastia de Reemplazo de Cadera
/
Polietileno
/
Prótesis de Cadera
Tipo de estudio:
Clinical_trials
/
Diagnostic_studies
/
Observational_studies
/
Prognostic_studies
Límite:
Adult
/
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma: