Estimating dose to implantable cardioverter-defibrillator outside the treatment fields using a skin QED diode, optically stimulated luminescent dosimeters, and LiF thermoluminescent dosimeters.
Chan, Maria F; Song, Yulin; Dauer, Lawrence T; Li, Jingdong; Huang, David; Burman, Chandra.
Afiliación
  • Chan MF; Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, NY, USA.
Med Dosim ; 37(3): 334-8, 2012.
Article en En | MEDLINE | ID: mdl-22296877
ABSTRACT
The purpose of this work was to determine the relative sensitivity of skin QED diodes, optically stimulated luminescent dosimeters (OSLDs) (microStar™ DOT, Landauer), and LiF thermoluminescent dosimeters (TLDs) as a function of distance from a photon beam field edge when applied to measure dose at out-of-field points. These detectors have been used to estimate radiation dose to patients' implantable cardioverter-defibrillators (ICDs) located outside the treatment field. The ICDs have a thin outer case made of 0.4- to 0.6-mm-thick titanium (∼2.4-mm tissue equivalent). A 5-mm bolus, being the equivalent depth of the devices under the patient's skin, was placed over the ICDs. Response per unit absorbed dose-to-water was measured for each of the dosimeters with and without bolus on the beam central axis (CAX) and at a distance up to 20 cm from the CAX. Doses were measured with an ionization chamber at various depths for 6- and 15-MV x-rays on a Varian Clinac-iX linear accelerator. Relative sensitivity of the detectors was determined as the ratio of the sensitivity at each off-axis distance to that at the CAX. The detector sensitivity as a function of the distance from the field edge changed by ± 3% (1-11%) for LiF TLD-700, decreased by 10% (5-21%) for OSLD, and increased by 16% (11-19%) for the skin QED diode (Sun Nuclear Corp.) at the equivalent depth of 5 mm for 6- or 15-MV photon energies. Our results showed that the use of bolus with proper thickness (i.e., ∼d(max) of the photon energy) on the top of the ICD would reduce the scattered dose to a lower level. Dosimeters should be calibrated out-of-field and preferably with bolus equal in thickness to the depth of interest. This can be readily performed in clinic.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Semiconductores / Fenómenos Fisiológicos de la Piel / Dosimetría Termoluminiscente / Desfibriladores Implantables / Radioterapia Conformacional / Dosimetría con Luminiscencia Ópticamente Estimulada Tipo de estudio: Diagnostic_studies Idioma: En Revista: Med Dosim Asunto de la revista: RADIOTERAPIA Año: 2012 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Semiconductores / Fenómenos Fisiológicos de la Piel / Dosimetría Termoluminiscente / Desfibriladores Implantables / Radioterapia Conformacional / Dosimetría con Luminiscencia Ópticamente Estimulada Tipo de estudio: Diagnostic_studies Idioma: En Revista: Med Dosim Asunto de la revista: RADIOTERAPIA Año: 2012 Tipo del documento: Article País de afiliación: Estados Unidos