Biological adaptation to misfits of immediately loaded fixed prostheses following computer-guided surgery.
J Prosthodont
; 21(3): 185-90, 2012 Apr.
Article
en En
| MEDLINE
| ID: mdl-22339733
In this clinical report, following computer-guided (3D Procera Software Planning Program, Nobel Biocare, Yorba Linda, CA) placement and immediate provisionalization of 12 dental implants (NobelSpeedy™ Replace, Nobel Biocare), misfits of the prefabricated screw-retained interim prostheses were noted at several implant-abutment junctions. Nevertheless, adaptation of the misfits was observed 10 days later, after the loosened screws were tightened. While a high mean marginal bone loss of 2.1 mm (range: 1.4 to 3.5 mm) was noted, all implants remained osseointegrated at 3-year follow-up.