Your browser doesn't support javascript.
loading
Reduction of thermal fluctuations in a cryogenic laser interferometric gravitational wave detector.
Uchiyama, Takashi; Miyoki, Shinji; Telada, Souichi; Yamamoto, Kazuhiro; Ohashi, Masatake; Agatsuma, Kazuhiro; Arai, Koji; Fujimoto, Masa-Katsu; Haruyama, Tomiyoshi; Kawamura, Seiji; Miyakawa, Osamu; Ohishi, Naoko; Saito, Takanori; Shintomi, Takakazu; Suzuki, Toshikazu; Takahashi, Ryutaro; Tatsumi, Daisuke.
Afiliación
  • Uchiyama T; Kamioka Observatory, Institute for Cosmic Ray Research, The University of Tokyo, 456 Higashi-Mozumi, Kamioka, Hida Gifu 506-1205, Japan. uchiyama@icrr.u-tokyo.ac.jp
Phys Rev Lett ; 108(14): 141101, 2012 Apr 06.
Article en En | MEDLINE | ID: mdl-22540781
ABSTRACT
The thermal fluctuation of mirror surfaces is the fundamental limitation for interferometric gravitational wave (GW) detectors. Here, we experimentally demonstrate for the first time a reduction in a mirror's thermal fluctuation in a GW detector with sapphire mirrors from the Cryogenic Laser Interferometer Observatory at 17 and 18 K. The detector sensitivity, which was limited by the mirror's thermal fluctuation at room temperature, was improved in the frequency range of 90 to 240 Hz by cooling the mirrors. The improved sensitivity reached a maximum of 2.2×10(-19) m/√Hz at 165 Hz.
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Phys Rev Lett Año: 2012 Tipo del documento: Article País de afiliación: Japón
Buscar en Google