Determination of phenylbutyric acid and its metabolite phenylacetic acid in different tissues of mouse by liquid chromatography with tandem mass spectrometry and its application in drug tissue distribution.
Marahatta, Anu; Bhandary, Bidur; Lee, Mi-Rin; Kim, Do-Sung; Lee, Yong Chul; Kim, So-Ri; Kim, Hyung-Ryong; Chae, Han-Jung.
Afiliación
  • Marahatta A; Department of Pharmacology, School of Medicine, Chonbuk National University, Jeonju 560-182, South Korea.
Article en En | MEDLINE | ID: mdl-22841743
ABSTRACT
Endoplasmic reticulum (ER) stress is associated with various human diseases. Phenylbutyric acid (PBA) is a well-known chemical chaperone that regulates ER stress. The main objective of this study was to develop a simple, rapid, and sensitive method for the simultaneous determination of phenylbutyric acid and its metabolite, phenylacetic acid (PAA). A LC-MS/MS analysis using negative electrospray ionization was used. Samples were analyzed by multiple reaction monitoring (MRM) in 15 min of total run time, using d11-PBA and d7-PAA as internal standards. The limit of quantification was 1 µg/g for tissue and 0.8 µg/mL for plasma. Recoveries for plasma and tissues were higher than 81% for both PBA and PAA. The inter-day and intra-day accuracy and precision were within ±15%. We then further successfully validated this method by applying it to determine the tissue distribution of PBA and its metabolite PAA after i.p. injection of PBA at a dose of 500 mg/kg in mice. The maximum concentrations of PBA and PAA in plasma and tissues were seen at 15 min and 45 min, respectively. The PBA plasma concentration was 15-fold higher than the concentration in the kidney, whereas the PAA plasma concentration was 6-fold higher than the concentration in the liver. The area under the curve decreased in the order of plasma > kidney > liver > heart > muscle > lung for PBA and plasma > liver > kidney > heart > muscle > lung for PAA. The tissue to plasma ratio ranged from 0.007 to 0.063 for PBA and 0.016 to 0.109 for PAA. In summary, the LC-ESI-MS method developed in this study is simple, sensitive and reliable.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fenilacetatos / Fenilbutiratos / Cromatografía Liquida / Espectrometría de Masas en Tándem Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2012 Tipo del documento: Article País de afiliación: Corea del Sur

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fenilacetatos / Fenilbutiratos / Cromatografía Liquida / Espectrometría de Masas en Tándem Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2012 Tipo del documento: Article País de afiliación: Corea del Sur