Wireless distributed functional electrical stimulation system.
Jovicic, Nenad S; Saranovac, Lazar V; Popovic, Dejan B.
Afiliación
  • Jovicic NS; School of Electrical Engineering, University of Belgrade, Belgrade, Serbia. nenad@etf.rs
J Neuroeng Rehabil ; 9: 54, 2012 Aug 09.
Article en En | MEDLINE | ID: mdl-22876934
ABSTRACT

BACKGROUND:

The control of movement in humans is hierarchical and distributed and uses feedback. An assistive system could be best integrated into the therapy of a human with a central nervous system lesion if the system is controlled in a similar manner. Here, we present a novel wireless architecture and routing protocol for a distributed functional electrical stimulation system that enables control of movement.

METHODS:

The new system comprises a set of miniature battery-powered devices with stimulating and sensing functionality mounted on the body of the subject. The devices communicate wirelessly with one coordinator device, which is connected to a host computer. The control algorithm runs on the computer in open- or closed-loop form. A prototype of the system was designed using commercial, off-the-shelf components. The propagation characteristics of electromagnetic waves and the distributed nature of the system were considered during the development of a two-hop routing protocol, which was implemented in the prototype's software.

RESULTS:

The outcomes of this research include a novel system architecture and routing protocol and a functional prototype based on commercial, off-the-shelf components. A proof-of-concept study was performed on a hemiplegic subject with paresis of the right arm. The subject was tasked with generating a fully functional palmar grasp (closing of the fingers). One node was used to provide this movement, while a second node controlled the activation of extensor muscles to eliminate undesired wrist flexion. The system was tested with the open- and closed-loop control algorithms.

CONCLUSIONS:

The system fulfilled technical and application requirements. The novel communication protocol enabled reliable real-time use of the system in both closed- and open-loop forms. The testing on a patient showed that the multi-node system could operate effectively to generate functional movement.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Terapia por Estimulación Eléctrica / Tecnología Inalámbrica Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: J Neuroeng Rehabil Asunto de la revista: ENGENHARIA BIOMEDICA / NEUROLOGIA / REABILITACAO Año: 2012 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Terapia por Estimulación Eléctrica / Tecnología Inalámbrica Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: J Neuroeng Rehabil Asunto de la revista: ENGENHARIA BIOMEDICA / NEUROLOGIA / REABILITACAO Año: 2012 Tipo del documento: Article