Prospective comparison of valve regurgitation quantitation by cardiac magnetic resonance imaging and transthoracic echocardiography.
Cawley, Peter J; Hamilton-Craig, Christian; Owens, David S; Krieger, Eric V; Strugnell, Wendy E; Mitsumori, Lee; D'Jang, Caryn L; Schwaegler, Rebecca G; Nguyen, Khanh Q; Nguyen, Bianca; Maki, Jeffrey H; Otto, Catherine M.
Afiliación
  • Cawley PJ; Division of Cardiology, Departments of Medicine, University of Washington School of Medicine, Seattle, WA, USA.
Circ Cardiovasc Imaging ; 6(1): 48-57, 2013 Jan 01.
Article en En | MEDLINE | ID: mdl-23212272
ABSTRACT

BACKGROUND:

Both transthoracic echocardiography (TTE) and cardiac magnetic resonance (CMR) imaging allow quantification of chronic aortic regurgitation (AR) and mitral regurgitation (MR). We hypothesized that CMR measurement of regurgitant volume (RVol) is more reproducible than TTE. METHODS AND

RESULTS:

TTE and CMR performed on the same day in 57 prospectively enrolled adults (31 with AR, 26 with MR) were measured by 2 independent physicians. TTE RVol(AR) was calculated as Doppler left ventricular outflow minus inflow stroke volume. RVol(MR) was calculated by both the proximal isovelocity surface area method and Doppler volume flow at 2 sites. CMR RVol(AR) was calculated by phase-contrast velocity mapping at the aortic sinuses and RVol(MR) as total left ventricular minus forward stroke volume. Intraobserver and interobserver variabilities were similar. For AR, the Bland-Altman mean interobserver difference in RVol was -0.7 mL (95% confidence interval [CI], -5 to 4) for CMR and -9 mL (95% CI, -53 to -36) for TTE. The Pearson correlation was higher (P=0.001) between CMR (0.99) than TTE readers (0.89). For MR, the Bland-Altman mean difference in RVol between observers was -4 mL (95% CI, -21 to 13) for CMR compared with 0.7 mL (95% CI, -30 to 32) for the proximal isovelocity surface area and -10 mL (95% CI, -76 to 56) for TTE volume flow at 2 sites. Correlation was similar for CMR (0.94) versus TTE readers (0.90 for the proximal isovelocity surface area).

CONCLUSIONS:

Compared with TTE, CMR has lower intraobserver and interobserver variabilities for RVol(AR), suggesting CMR may be superior for serial measurements. Although RVol(MR) is similar by TTE and CMR, variability in measured RVol by both approaches suggests that caution is needed in clinical practice.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Insuficiencia de la Válvula Aórtica / Ecocardiografía Doppler / Función Ventricular Izquierda / Imagen por Resonancia Cinemagnética / Insuficiencia de la Válvula Mitral Tipo de estudio: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Circ Cardiovasc Imaging Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Insuficiencia de la Válvula Aórtica / Ecocardiografía Doppler / Función Ventricular Izquierda / Imagen por Resonancia Cinemagnética / Insuficiencia de la Válvula Mitral Tipo de estudio: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Circ Cardiovasc Imaging Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos
...