Knee wear simulation under conditions of highly demanding daily activities--influence on an unicompartmental fixed bearing knee design.
Schwiesau, Jens; Schilling, Carolin; Utzschneider, Sandra; Jansson, Volkmar; Fritz, Bernhard; Blömer, Wilhelm; Grupp, Thomas M.
Afiliación
  • Schwiesau J; Aesculap AG, Research & Development, Tuttlingen, Germany.
Med Eng Phys ; 35(8): 1204-11, 2013 Aug.
Article en En | MEDLINE | ID: mdl-23380535
ABSTRACT
The objectives of our in vitro study were to evaluate a knee wear simulation based on patient daily activities in combination with artificial ageing of polyethylene inserts to create an optimised simulation of in vivo wear modes. A wear simulation was performed on fixed bearing unicompartmental knee arthroplasty (UKA) devices in a direct comparison of level walking (as given by the ISO 14243-12002(E) profiles) and in a customised test configuration based on activities for level walking (10%), stairs ascending (40%), stairs descending (40%), chair rising (8%) and deep squatting (2%). The cumulative gravimetric wear was estimated to be 15.3mg for level walking (ISO) and 69.6 mg for high demanding activities (HDA). The gravimetric wear rate of the ISO group was 3.0mg/million cycles, compared to 11.7 mg/million cycles for the HDA protocol. Level walking wear testing conditions (ISO) and artificial ageing alone is not sufficient to reproduce in vivo failure modes. After 3 million cycles all gliding surfaces of the HDA group developed in the tibio-femoral articulation markable areas of structural material fatigue and delamination. In conclusion a combination of artificial ageing to clinical relevant oxidation grades and a sequence of various high demanding daily patient activities is necessary to represent a revised in vitro behaviour of abrasive-adhesive wear and delamination in artificial knee replacements.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Actividades Cotidianas / Análisis de Falla de Equipo / Polietileno / Biomimética / Articulación de la Rodilla / Prótesis de la Rodilla / Actividad Motora Tipo de estudio: Evaluation_studies / Guideline Límite: Humans Idioma: En Revista: Med Eng Phys Asunto de la revista: BIOFISICA / ENGENHARIA BIOMEDICA Año: 2013 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Actividades Cotidianas / Análisis de Falla de Equipo / Polietileno / Biomimética / Articulación de la Rodilla / Prótesis de la Rodilla / Actividad Motora Tipo de estudio: Evaluation_studies / Guideline Límite: Humans Idioma: En Revista: Med Eng Phys Asunto de la revista: BIOFISICA / ENGENHARIA BIOMEDICA Año: 2013 Tipo del documento: Article País de afiliación: Alemania