Your browser doesn't support javascript.
loading
SCOUT: a fast Monte-Carlo modeling tool of scintillation camera output.
Hunter, William C J; Barrett, Harrison H; Muzi, John P; McDougald, Wendy; MacDonald, Lawrence R; Miyaoka, Robert S; Lewellen, Thomas K.
Afiliación
  • Hunter WC; Department of Radiology, University of Washington, Seattle, WA 98195, USA. wcjh@uw.edu
Phys Med Biol ; 58(11): 3581-98, 2013 Jun 07.
Article en En | MEDLINE | ID: mdl-23640136
ABSTRACT
We have developed a Monte-Carlo photon-tracking and readout simulator called SCOUT to study the stochastic behavior of signals output from a simplified rectangular scintillation-camera design. SCOUT models the salient processes affecting signal generation, transport, and readout of a scintillation camera. In this work, we compare output signal statistics from SCOUT to experimental results for both a discrete and a monolithic camera. We also benchmark the speed of this simulation tool and compare it to existing simulation tools. We find this modeling tool to be relatively fast and predictive of experimental results. Depending on the modeled camera geometry, we found SCOUT to be 4 to 140 times faster than other modeling tools.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Conteo por Cintilación / Método de Montecarlo Tipo de estudio: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Phys Med Biol Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Conteo por Cintilación / Método de Montecarlo Tipo de estudio: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Phys Med Biol Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos