Last-recorded P2Y12 reaction units value is strongly associated with thromboembolic and hemorrhagic complications occurring up to 6 months after treatment in patients with cerebral aneurysms treated with the pipeline embolization device.
AJNR Am J Neuroradiol
; 35(1): 128-35, 2014 Jan.
Article
en En
| MEDLINE
| ID: mdl-23828107
BACKGROUND AND PURPOSE: A recent study identified a preprocedural P2Y12 reaction units value of <60 or >240 as a strong independent predictor of perioperative thromboembolic and hemorrhagic complications after treatment of cerebral aneurysms with the Pipeline Embolization Device. This study aimed to determine whether a last-recorded P2Y12 reaction units value of <60 or >240 predicts thromboembolic and hemorrhagic complications up to 6 months after treatment of cerebral aneurysms with the Pipeline Embolization Device in the same patient cohort. MATERIALS AND METHODS: We recorded patient and aneurysm characteristics, P2Y12 receptor antagonist administered, P2Y12 reaction units value with VerifyNow, procedural variables, and thromboembolic and hemorrhagic complications up to 6 months after Pipeline Embolization Device procedures at our institution during an 8-month period. Complications causing a permanent disabling neurologic deficit or death were considered major. Multivariate regression analysis was performed to identify independent predictors of thromboembolic and hemorrhagic complications. RESULTS: Forty-four patients underwent 48 Pipeline Embolization Device procedures at our institution during the study period. There were 11 thromboembolic and hemorrhagic complications up to 6 months after treatment in our cohort (22.9%), 5 of which were major (10.4%). A last-recorded P2Y12 reaction units value of <60 or >240 was the only independent predictor of all (P = .002) and major (P = .03) thromboembolic and hemorrhagic complications in our cohort. Most patients (71%) required, on average, 2 adjustments to the dose or type of P2Y12 receptor antagonist to remain within the 60-240 target P2Y12 reaction units range. CONCLUSIONS: In our cohort, a last-recorded P2Y12 reaction units value of <60 or >240 was the only independent predictor of all and major thromboembolic and hemorrhagic complications up to 6 months after Pipeline Embolization Device procedures.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Aneurisma Intracraneal
/
Hemorragia Cerebral
/
Embolia Intracraneal
/
Trombosis Intracraneal
/
Embolización Terapéutica
/
Receptores Purinérgicos P2Y12
Tipo de estudio:
Diagnostic_studies
/
Etiology_studies
/
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Límite:
Adult
/
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
AJNR Am J Neuroradiol
Año:
2014
Tipo del documento:
Article
Pais de publicación: