Accelerated polyethylene wear and early failure of the uncemented Reflection acetabular component: a ten to fifteen-year follow-up study.
Clement, Nicholas D; Mitchell, Martin; Trayner, Morag E; Porter, Daniel E; Lawson, Graham M; Burnett, Richard.
Afiliación
  • Clement ND; Department of Orthopaedics and Trauma, The Royal Infirmary of Edinburgh, Little France, Edinburgh, EH16 4SA, UK, nickclement@doctors.org.uk.
Int Orthop ; 37(12): 2337-43, 2013 Dec.
Article en En | MEDLINE | ID: mdl-23974837
PURPOSE: The Reflection® second-generation uncemented acetabular component was designed to address increased rates of failure observed with the early acetabular designs. However, the reported survivorship of this acetabular component has been conflicting. The aim of this study was to describe the ten to 15-year survivorship and polyethylene wear rate for the uncemented Reflection® acetabular component performed as part of a primary total hip replacement. METHODS: One hundred and four consecutive Reflection® uncemented acetabular components in 97 patients were identified from a prospective arthroplasty database with a minimum of ten years of follow-up. No patient was lost to follow-up. Mean cohort age was 59.1 years. RESULTS: There were 24 revisions of the acetabular component. The all-cause survival rate at ten years was 77.2% [95% confidence interval (CI) 73.9-80.5]. Mean linear wear was 0.20 mm [standard deviation (SD) 0.08] per year and the volumetric wear was 106.2 mm(3) per year. At final follow-up, more than half the patients had osteolysis identified around the femoral component. The mean Oxford Hip Score was 35.6 (SD 9.8) at a mean follow-up of 11.9 years, with six (11.1%) excellent, 26 (48.1%) good, 17 (31.5%) fair and five (9.3%) poor outcomes. Despite the fact that more than a third had a fair or poor outcome, only four (7.4%) were not satisfied with their hip. CONCLUSIONS: Due to the high rate of relatively asymptomatic polyethylene wear and osteolysis associated with this acetabular component, in our department, we now review all surviving patients both clinically and radiographically on an annual basis.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Diseño de Prótesis / Falla de Prótesis / Osteoartritis de la Cadera / Artroplastia de Reemplazo de Cadera / Polietileno / Prótesis de Cadera / Acetábulo Tipo de estudio: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Año: 2013 Tipo del documento: Article Pais de publicación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Diseño de Prótesis / Falla de Prótesis / Osteoartritis de la Cadera / Artroplastia de Reemplazo de Cadera / Polietileno / Prótesis de Cadera / Acetábulo Tipo de estudio: Diagnostic_studies / Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Año: 2013 Tipo del documento: Article Pais de publicación: Alemania