Simultaneous dual wavelength eye-tracked ultrahigh resolution retinal and choroidal optical coherence tomography.
Opt Lett ; 38(21): 4312-5, 2013 Nov 01.
Article en En | MEDLINE | ID: mdl-24177081
ABSTRACT
We demonstrate an optical coherence tomography device that simultaneously combines different novel ultrabroad bandwidth light sources centered in the 800 and 1060 nm regions, operating at 66 kHz depth scan rate, and a confocal laser scanning ophthalmoscope-based eye tracker to permit motion-artifact-free, ultrahigh resolution and high contrast retinal and choroidal imaging. The two wavelengths of the device provide the complementary information needed for diagnosis of subtle retinal changes, while also increasing visibility of deeper-lying layers to image pathologies that include opaque media in the anterior eye segment or eyes with increased choroidal thickness.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Retina / Coroides / Tomografía de Coherencia Óptica / Movimientos Oculares Límite: Humans Idioma: En Revista: Opt Lett Año: 2013 Tipo del documento: Article