Use of a Si(Li) detector as ß spectrometer.
Dryák, P; Kovár, P.
Afiliación
  • Dryák P; Czech Metrology Institute, Radiova 1, CZ-102 00 Prague 10, Czech Republic. Electronic address: pdryak@cmi.cz.
  • Kovár P; Czech Metrology Institute, Radiova 1, CZ-102 00 Prague 10, Czech Republic.
Appl Radiat Isot ; 87: 325-7, 2014 May.
Article en En | MEDLINE | ID: mdl-24291526
ABSTRACT
The aim of this work is to demonstrate the capability of a Si(Li) detector for the measurement of ß spectra, despite the energy absorption in air and in the Be window. A simple source holder fixes the source on the symmetry axis at 3mm from the detector window. The ß-sources are produced by evaporation on a plastic backing plate. Absorbing materials between the source and the sensitive volume of the detector are 3 mm of air, a Be window, 0.1 µm Si and 20 nm of gold. A model of the detector was created for ß spectra simulation using the MCNP 4A code. Experimental spectra of (14)C, (147)Pm, (204)Tl, (90)Sr/(90)Y were compared with simulated spectra.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2014 Tipo del documento: Article Pais de publicación: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2014 Tipo del documento: Article Pais de publicación: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM