Combined acquisition technique (CAT) for neuroimaging of multiple sclerosis at low specific absorption rates (SAR).
Biller, Armin; Choli, Morwan; Blaimer, Martin; Breuer, Felix A; Jakob, Peter M; Bartsch, Andreas J.
Afiliación
  • Biller A; Department of Neuroradiology, University of Heidelberg, Heidelberg, Germany.
  • Choli M; Research Center for Magnetic-Resonance-Bavaria (MRB), Wuerzburg, Germany.
  • Blaimer M; Research Center for Magnetic-Resonance-Bavaria (MRB), Wuerzburg, Germany.
  • Breuer FA; Research Center for Magnetic-Resonance-Bavaria (MRB), Wuerzburg, Germany.
  • Jakob PM; Research Center for Magnetic-Resonance-Bavaria (MRB), Wuerzburg, Germany; Department of Experimental Physics 5, University of Wuerzburg, Wuerzburg, Germany.
  • Bartsch AJ; Department of Neuroradiology, University of Heidelberg, Heidelberg, Germany; FMRIB Centre, University of Oxford, Oxford, United Kingdom.
PLoS One ; 9(3): e91030, 2014.
Article en En | MEDLINE | ID: mdl-24608106
ABSTRACT

PURPOSE:

To compare a novel combined acquisition technique (CAT) of turbo-spin-echo (TSE) and echo-planar-imaging (EPI) with conventional TSE. CAT reduces the electromagnetic energy load transmitted for spin excitation. This radiofrequency (RF) burden is limited by the specific absorption rate (SAR) for patient safety. SAR limits restrict high-field MRI applications, in particular. MATERIAL AND

METHODS:

The study was approved by the local Medical Ethics Committee. Written informed consent was obtained from all participants. T2- and PD-weighted brain images of n = 40 Multiple Sclerosis (MS) patients were acquired by CAT and TSE at 3 Tesla. Lesions were recorded by two blinded, board-certificated neuroradiologists. Diagnostic equivalence of CAT and TSE to detect MS lesions was evaluated along with their SAR, sound pressure level (SPL) and sensations of acoustic noise, heating, vibration and peripheral nerve stimulation.

RESULTS:

Every MS lesion revealed on TSE was detected by CAT according to both raters (Cohen's kappa of within-rater/across-CAT/TSE lesion detection κCAT = 1.00, at an inter-rater lesion detection agreement of κLES = 0.82). CAT reduced the SAR burden significantly compared to TSE (p<0.001). Mean SAR differences between TSE and CAT were 29.0 (± 5.7) % for the T2-contrast and 32.7 (± 21.9) % for the PD-contrast (expressed as percentages of the effective SAR limit of 3.2 W/kg for head examinations). Average SPL of CAT was no louder than during TSE. Sensations of CAT- vs. TSE-induced heating, noise and scanning vibrations did not differ.

CONCLUSION:

T2-/PD-CAT is diagnostically equivalent to TSE for MS lesion detection yet substantially reduces the RF exposure. Such SAR reduction facilitates high-field MRI applications at 3 Tesla or above and corresponding protocol standardizations but CAT can also be used to scan faster, at higher resolution or with more slices. According to our data, CAT is no more uncomfortable than TSE scanning.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen Eco-Planar / Neuroimagen / Esclerosis Múltiple Tipo de estudio: Guideline Aspecto: Ethics Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2014 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen Eco-Planar / Neuroimagen / Esclerosis Múltiple Tipo de estudio: Guideline Aspecto: Ethics Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2014 Tipo del documento: Article País de afiliación: Alemania