Your browser doesn't support javascript.
loading
Optical fiber Sagnac interferometer for sensing scalar directional refraction: application to magnetochiral birefringence.
Loas, G; Alouini, M; Vallet, M.
Afiliación
  • Loas G; Institut de Physique de Rennes, UMR 6251, CNRS, Université de Rennes 1 Campus de Beaulieu, Rennes 35042, France.
  • Alouini M; Institut de Physique de Rennes, UMR 6251, CNRS, Université de Rennes 1 Campus de Beaulieu, Rennes 35042, France.
  • Vallet M; Institut de Physique de Rennes, UMR 6251, CNRS, Université de Rennes 1 Campus de Beaulieu, Rennes 35042, France.
Rev Sci Instrum ; 85(4): 043109, 2014 Apr.
Article en En | MEDLINE | ID: mdl-24784595
ABSTRACT
We present a setup dedicated to the measurement of the small scalar directional anisotropies associated to the magnetochiral interaction. The apparatus, based on a polarization-independent fiber Sagnac interferometer, is optimized to be insensitive to circular anisotropies and to residual absorption. It can thus characterize samples of biological interests, for which the two enantiomers are not available and/or which present poor transmission. The signal-to-noise ratio is shown to be limited only by the source intensity noise, leading to a detection limit of Δϕ = 500 nrad Hz(-1/2). It yields a limit on the magnetochiral index nMC < 4 × 10(-13) T(-1) at 1.55 µm for the organic molecules tested.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2014 Tipo del documento: Article País de afiliación: Francia