Your browser doesn't support javascript.
loading
New spectral-spatial imaging algorithm for full EPR spectra of multiline nitroxides and pH sensitive trityl radicals.
Tseitlin, Mark; Biller, Joshua R; Elajaili, Hanan; Khramtsov, Valery V; Dhimitruka, Ilirian; Eaton, Gareth R; Eaton, Sandra S.
Afiliación
  • Tseitlin M; Department of Chemistry and Biochemistry, University of Denver, Denver, CO 80210, USA.
  • Biller JR; Department of Chemistry and Biochemistry, University of Denver, Denver, CO 80210, USA.
  • Elajaili H; Department of Chemistry and Biochemistry, University of Denver, Denver, CO 80210, USA.
  • Khramtsov VV; Dorothy M. Davis Heart & Lung Research Institute and Division of Pulmonary, Allergy, Critical Care & Sleep Medicine, Department of Internal Medicine, The Ohio State University, Columbus, OH 43210, USA.
  • Dhimitruka I; Dorothy M. Davis Heart & Lung Research Institute and Division of Pulmonary, Allergy, Critical Care & Sleep Medicine, Department of Internal Medicine, The Ohio State University, Columbus, OH 43210, USA.
  • Eaton GR; Department of Chemistry and Biochemistry, University of Denver, Denver, CO 80210, USA. Electronic address: geaton@du.edu.
  • Eaton SS; Department of Chemistry and Biochemistry, University of Denver, Denver, CO 80210, USA.
J Magn Reson ; 245: 150-5, 2014 Aug.
Article en En | MEDLINE | ID: mdl-25058914
ABSTRACT
An algorithm is derived and demonstrated that reconstructs an EPR spectral-spatial image from projections with arbitrarily selected gradients. This approach permits imaging wide spectra without the use of the very large sweep widths and gradients that would be required for spectral-spatial imaging with filtered back projection reconstruction. Each projection is defined as the sum of contributions at the set of locations in the object. At each location gradients shift the spectra in the magnetic field domain, which is equivalent to a phase change in the Fourier-conjugate frequency domain. This permits solution of the problem in the frequency domain. The method was demonstrated for 2D images of phantoms consisting of (i) two tubes containing (14)N and (15)N nitroxide and (ii) two tubes containing a pH sensitive trityl radical at pH 7.0 and 7.2. In each case spectral slices through the image agree well with the full spectra obtained in the absence of gradient.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Compuestos de Tritilo / Algoritmos / Espectroscopía de Resonancia por Spin del Electrón / Óxidos de Nitrógeno Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Revista: J Magn Reson Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2014 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Compuestos de Tritilo / Algoritmos / Espectroscopía de Resonancia por Spin del Electrón / Óxidos de Nitrógeno Tipo de estudio: Diagnostic_studies / Prognostic_studies Idioma: En Revista: J Magn Reson Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2014 Tipo del documento: Article País de afiliación: Estados Unidos