The effect of application time of two types of bone cement on the cement-bone interface strength.
Dahabreh, Ziad; Phillips, Hannah Kalpana; Stewart, Todd; Stone, Martin.
Afiliación
  • Dahabreh Z; Department of Trauma and Orthopaedics, Leeds General Infirmary, Great George Street, Leeds, LS1 3EX, UK.
Eur J Orthop Surg Traumatol ; 25(4): 775-81, 2015 May.
Article en En | MEDLINE | ID: mdl-25192848
The aim of this study was to investigate whether the application time of bone cement would have an effect on the cement-bone interface strength in two types of commercially available bone cements. CMW1 Radiopaque(®) (CMW1) and SmartSetHV(®) (SmartSet) were applied to bovine cancellous bone specimens at 2 and at 4 min. Specimens were loaded to failure and the shear strength of the cement-bone interface was calculated. The mean shear strength (±standard deviation) of the cement-bone interface was 2.79 ± 1.29 MPa for CMW1 applied at 2 min; 1.35 ± 0.89 MPa for CMW1 applied at 4 min; 2.93 ± 1.21 MPa for SmartSet applied at 2 min and 3.00 ± 1.11 MPa for SmartSet applied at 4 min. Compared to all other groups, the cement-bone interface strength was significantly lower when CMW1 was applied to the bone specimens at 4 min (p < 0.05). There was no significant difference in the cement-bone interface strength when SmartSet was applied to bone at 2 and at 4 min. Under these testing conditions, the cement-bone interface strength was not affected by the time of application of SmartSet to bone. However, it was significantly lower when CMW1 was applied to bone at 4 min.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tibia / Cementos para Huesos / Polimetil Metacrilato / Interfase Hueso-Implante Límite: Animals Idioma: En Revista: Eur J Orthop Surg Traumatol Año: 2015 Tipo del documento: Article Pais de publicación: Francia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tibia / Cementos para Huesos / Polimetil Metacrilato / Interfase Hueso-Implante Límite: Animals Idioma: En Revista: Eur J Orthop Surg Traumatol Año: 2015 Tipo del documento: Article Pais de publicación: Francia